α β γ ray Radiation Room

In the α β γ room are calibrated instruments for non Accredia report.

Liquid and extended sources are present in the room, used for calibration of the instruments used in nuclear medicine.

  • The extended sources are used to calibrate contamination meters Hands, Feet and Clothing meters
  • The liquid sources are used to calibrate the Dose calibrators.

The Comecer Calibration Report is issued following calibration, relative to the quantity of the “Activity”.

Electro-deposited sources for radioactive surface contamination

For the control of radioactive surface contamination of the instruments to be calibrated, currently the following reference radioactive sources are available:

Extended sources dim. 10×10 cm:

  • Source of Am241 (init. nominal activity: kBq 2.69)
  • Source of C14 (init. nominal activity: kBq 2.88)
  • Source of I129 (init. nominal activity: kBq 30)
  • Source of Cs137 (init. nominal activity: kBq 12)
  • Source of Sr90 (init. nominal activity: kBq 3.00)
  • Source of Co57 (init. nominal activity: kBq 35)
  • Source of Tc99 (init. nominal activity: kBq 2.32)

Sources vial for calibration Dose calibrators

For the calibration of Dose calibrators, the following sources are available sealed in 20 ml vials

  • Source of Cs137 (init. nominal activity: kBq 10000)
  • Source of Co60 (init. nominal activity: kBq 3645)
  • Source of Na22 (init. nominal activity: kBq 4180)